Model 572 Pressure Transmitter

Brand: Delta Controls Corporation

The Delta Controls Model 572 Level and Pressure Transmitter was designed specifically for food, pharmaceutical and ultraclean applications. 

The 572 consists of a potted electronic module in a housing which is mounted with a modular pressure sensor as a packaged assembly. 

The wetted sensor is constructed of 316 S.S. The body is polished and free from crevices, cracks, and voids that can harbor contaminants. 

The back of the sensor is usually vented to atmosphere to compensate for barometric pressure changes. 

Cleaning of the Delta Controls Model 572 can easily be accomplished by disassembly and manual washing, or it can be cleaned and sterilized in place. 

The Model 572 can also be used as a liquid level transmitter in vented tanks. The pressure at the bottom of a vented tank containing liquid is linearly proportional to the height and density of the liquid. Use the TPC Corrector to compensate for the gas space pressure when the tank is pressurized.
Features
  • Two wire loop-powered 4-20 mADC signal
  • Fits Ladish, APC, and other sanitary connections
  • Suitable for ultraclean water service
  • Suitable for CIP/SIP service
  • Removable electronic module
  • Polished body is basic, polished housing optional
  • Zero & span pots for field recalibration
  • Spark & power surge protection
  • Optional dampening
  • Optional LCD display
  • Maintenance free operation
  • Dampening: Selectable 0.1 to 22 seconds
  • Indicator: Large direct reading LCD
  • Alarm Points: 1 or 2, 5 Amp SPDT Contacts
  • Power Supply: Integral Vac to 24 Vdc
  • TPC Corrector: Level measurement in unvented and pressurized tanks. (Use for gas space compensation inpressurized tanks)

Specifications

Accuracy

  • ± 0.25 % of Full Scale or Better

Adjustments

  • Span from 30 % to 100 % of Range, Zero to 30 % is Basic, Non-interactive.

Barometric Effect

  • None

Loop Impedance

  • 550 ohms at 24 Vdc,1100 ohms at 35 Vdc

Output Signal

  • 4-20 mADC, Isolated, Intrinsically Safe

Over Pressure

  • 2x Range, 35 psi (2.5 bar) Min. without Damage, 3x Range 1000 psi (70 bar) Min. without Rupture

Supply Power

  • 13 to 35 Vdc 2-Wire Loop-Powered

Technology

  • Silicon Strain Gage

Temperature Range

  • -20° to +185° F (-30° to +85° C) Electronics
  • -20° to +220° F (-30° to 105° C) Process
  • 30° to +130° F (-1° to +55° C) Compensated (Higher Available)

Thermal Error

  • ± 0.02 % of Full Scale/°F Max.